Variation in Hemoglobin A1c Linked to Hemoglobin S: Comparison of Analysis Results from ion Exchange Chromatography and an Immunoturbidimetric Method

Niokhor Ndane Diouf, O. Boye, M. N. Coly, M. B. Touré, M. Djité, N. F. Coly, I. Y. Soumah, F. Diedhiou, A. Samba, I. Faye, M. M. Gueye, H. H. M. Agossou, R. Ba, S. Thiam, A. Ndiaye, F. Cisse, F. Diallo, M. Sylla-Niang, A. Diatta

Asian Journal of Biochemistry, Genetics and Molecular Biology · pp. 1–7 · Published 12 Dec 2023

10.9734/ajbgmb/2023/v15i4346

Abstract

Hemoglobin S can interfere with the measurement of glycated hemoglobin, an essential tool for diagnosing and monitoring diabetes. The objective of this study is to evaluate the analytical performance of  2 glycated haemoglobin assay methods. A prospective cross-sectional study was conducted where 186 patients (61 homozygous sickle cell disease, 61 AA type subjects and 64 AS type subjects) were recruited. Glycated hemoglobin was measured by immunoturbidimetry method and ion exchange chromatography. The coefficient of variation (CV) of the repeatability is 2.34% and 1.13% (normal rate) ; 2.32% and 1.65% (high rate) respectively for the immunoturbidimetric method and ion exchange chromatography. In reproducibility, the CV obtained are 3.23% and 2.64% (normal rate) and 3.27% and 2.22% (high rate), respectively for the immunoturbidimetric method and the ion exchange chromatography. Linearity is satisfactory for both methods. Mean glycated hemoglobin values ​​show no significant difference (the P-value is equal to 0.09, 0.17 and 0.70 respectively in subjects AA, AS and SS) in the 2 methods for patients with the same hemoglobin electrophoretic profile. The analytical performances of the 2 methods are good but their use is not recommended in the biological diagnosis of diabetes and pre-diabetes due to interference from hemoglobin S, especially in the case of homozygous sickle cell disease or in the case of composite heterozygosity.

Hemoglobin S glycated hemoglobin immunoturbidimetric method ion exchange chromatography

Cited by 0

No indexed citations yet.

Article metrics

Real usage data collected on this platform.

0

Page views

0

PDF downloads

0

Outbound clicks

0

Citations

Views by country

Approximate, from request IP at view time — not citizenship or institution. Countries with fewer than 5 views are grouped as "Other".

No views recorded yet.

Traffic sources

Referring site, by host.

No traffic recorded yet.

Views and downloads exclude known bots/crawlers. Citations combines this platform's own DOI-resolved index with each external source's own reported total — see Cited by above for individually listed citing works. Last refreshed 0 seconds ago.