Reaction Force Measurement in Milwaukee Brace during Daily Activities-Introduction of a New Measurement System

Ehsan Gohari, Ali Ataei, Mohammad Haghpanahi, Mohammadreza Mallakzadeh, Mohammad Parnianpour, Mohammadsaleh Ganjavian, Mojtaba Kamyab, Arezoo Amirpourabasi, Mohammad Saleh Khajeh Hosseini

Annual Research & Review in Biology · pp. 2856–2866 · Published 12 May 2014

10.9734/ARRB/2014/7588

Abstract

Aims: The reaction forces of the Milwaukee Brace’s (MB’s) several areas were measured during normal daily activities (24 hr) by a new measurement device improved in order to force acquisition by groups of sensors distributed in all of the MB’s internal areas. Methodology: In the current study, a new measurement system using Force Sensitive Resistance (FSRs) distributed on the different areas of MB for scoliosis patients. Sensitivity analysis has been fulfilled to compare two different sensors arrangements placed in the basket of MB. A reliability analysis has been carried out for normal activities including sitting, walking and sleeping. Two healthy volunteers and two scoliosis patients were nominated to force acquisition from their MB during 23 hours of daily activities. Results: In the sensitivity analysis, maximum value of Cronbach’s Alpha was 0.775 for sitting at basket lateral side. In the reliability analysis, the maximum and minimum difference between the force values of the two arrangements is 1.37(N) and 0.19(N) respectively related to the sitting test. The tests for each subject commenced with 9-11 hours of sleeping and continued for 12-14 hours in which main activities were sitting and walking. The forces average values of each part of the tests were reported for different areas of MBs. The mean, median, mode, standard deviation and ratio of standard deviation on the mean have been calculated for each brace region for both time durations. Conclusion: The brace different areas forces for sleeping and day time activities had normal distribution. The force value of the brace regions having direct contact with mattress surface was increased during sleeping with respect to day time activities. Other regions forces were decreased or remained constant in this time duration. The forces detected from the lumbar pad during most of the test time were constant.

Adolescent idiopathic scoliosis milwaukee brace force measuring clinical test.

Cited by 0

No indexed citations yet.

Article metrics

Real usage data collected on this platform.

0

Page views

0

PDF downloads

0

Outbound clicks

0

Citations

Views by country

Approximate, from request IP at view time — not citizenship or institution. Countries with fewer than 5 views are grouped as "Other".

No views recorded yet.

Traffic sources

Referring site, by host.

No traffic recorded yet.

Views and downloads exclude known bots/crawlers. Citations combines this platform's own DOI-resolved index with each external source's own reported total — see Cited by above for individually listed citing works. Last refreshed 0 seconds ago.