Ex vivo Analysis of Three Electronic Apex Locators Accuracy with Different Settings
Antonio Batista, Gilson Blitzkow Sydney, Marili Doro Andrade Deonizio, Thais Cristina Pereira, Carolina Carvalho de Oliveira Santos, Bruno Cavalini Cavenago
Journal of Advances in Medicine and Medical Research · pp. 1–7 · Published 14 Jul 2016
10.9734/BJMMR/2016/26477Abstract
Aims: The correct determination of a root canal length is a fundamental step for an adequate chemical-mechanical preparation and consequently, for a successful endodontic treatment. Electronic apex locators (EAL) were developed to provide fast and reliable working lengths. The aim of the present study was to analyze the accuracy of the Root ZX II, RomiApex A-15, and SmarPex EAL’s to determine the location of apical constriction, with and without the instructions recommended by the manufacturers. Methodology: Fifteen mandibular premolars were randomly selected and root canals were accessed. The real canal length was determined by introducing a #15 K-file until the tip was visualized in the apical foramen, using 40x magnification of an operative microscope. In the sequence, the teeth were inserted in plastic flasks containing floral foam soaked in 0.9% saline solution. The root canals were filled with 1% sodium hypochlorite and the electronic measures were obtained with the selected devices until the “0.0” or the last green bar mark, as showed in devices display, and as per the manufacturers settings recommendation. The data were submitted to statistical analysis with the Friedman and Wilcoxon tests with a 0.05% significance level (p < 0.05). Results: All devices were similar (p > 0.05) and showed precise and acceptable measurements at both times. Without manufacturers setting recommendation, the Root ZX II was the EAL that presented the greater percentage of coincidences with the real teeth length measures (73.33%), followed by the RomiApex A-15 (66.66%) and the SmarPex (40%). After performing the recommended settings the Root ZX II and the SmarPex presented 86.66% of coincidence with the real length, however, only the SmarPex device enhanced the mean precision with the real length (p < 0.05). Conclusion: Considering a clinically acceptable average error of ±0.5 mm, all devices were effective in determining the measurement until the apical constriction. Although, there were no statistical significant difference with and without manufacturers instruction, for Root ZX II, the performance was better when manufacturer instructions were followed.
Cited by 0
No indexed citations yet.
Related research
- Effectiveness of Cordia verbenacea and Propolis as Irrigation Solution on the Bond Strength of Filling Material to the Root Dentin: In vitro Study — shares topic coverage
- Low Intensity Laser Therapy: A Review of Its Applications and Effectiveness in Health, General Dentistry and Endodontics — shares topic coverage
- Salvaging an Endodontically Treated Mandibular Molar with a Furcation Perforation by Root Hemisection Followed by Fixed Prosthesis: A Case Report — shares topic coverage
- Influence of Plastic Endodontic Instruments on the Activation Protocol of Irrigantic Solutions in Chemomechanical Preparation: Literature Review — shares topic coverage
- Would the Accuracy of Dental Apex Locators in Determining the Apical Constriction be Influenced by the Activation Sequence? An Ex vivo Study — shares topic coverage
Article metrics
Real usage data collected on this platform.
0
Page views
0
PDF downloads
0
Outbound clicks
0
Citations
Views by country
Approximate, from request IP at view time — not citizenship or institution. Countries with fewer than 5 views are grouped as "Other".
No views recorded yet.
Traffic sources
Referring site, by host.
No traffic recorded yet.
Views and downloads exclude known bots/crawlers. Citations combines this platform's own DOI-resolved index with each external source's own reported total — see Cited by above for individually listed citing works. Last refreshed 0 seconds ago.