Dosimetric Characterizations of Megavoltage Therapeutic Photon Beam

Md. Abdullah Al Mashud, M. Jahangir Alam

Annual Research & Review in Biology · pp. 83–88 · Published 2 Sep 2020

10.9734/arrb/2020/v35i930280

Abstract

This paper presents the dosimetric parameters characterizations of a megavoltage therapeutic photon beam. The main focus of this study is to investigate and analyze the parameters of percentage depth dose (PDD) and tissue maximum ratio (TMR) due to the importance of treatment system. The depth dose characteristics of 6MV photon beam for different field sizes in water phantom has been measured, analyzed and found a robustness results. The results revealed that the depth dose variation from 0.067% to 1.812% and the TMR values varies from 0.501% to 2.111%. It seems the measured dosimetric quantities are clinically relevant for different field sizes and depths.

PDD TMR accuracy LINAC photon beam

References (14)

  1. 1 Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry based on Standards of Absorbed Dose to Water
  2. 2 Absorbed Dose Determination in External Beam Radiotherapy [DOI]
  3. 3 Dose calculation algorithms in external beam photon radiation therapy [DOI]
  4. 4 What degree of accuracy is required and can be achieved in photon and neutron therapy? [DOI]
  5. 5 Addendum to the IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al–4 mm Cu HVL) [DOI]
  6. 6 Estimation of inhomogenity correction factors for a Co-60 beam using Monte Carlo simulation [DOI]
  7. 7 Evaluation of depth dose characteristics of superficial X-rays machine using different kVp and applicators diameter
  8. 8 A filtration method for improving film dosimetry in photon radiation therapy [DOI]
  9. 9 Calculation of Analytical Expressions for Measured Percentage Depth Dose Data in Megavoltage Photon Therapy
  10. 10 The comparison of absorbed dose measurements for water and artificial body fluid
  11. 11 Analysis and Verification of Percent Depth Dose and Tissue Maximum Ratio for Co-60 Gamma Ray Beam
  12. 12 Dosimetric Verification of Bhabhatron - II Telecobalt Unit at the University College Hospital, Ibadan, Nigeria.
  13. 13 Percentage Discrepancies Assessment Between Measured and Calculated Behaviour of Percent Depth Dose in External Beam Radiotherapy [DOI]
  14. 14 International Journal of Cancer Therapy and Oncology [DOI]

Cited by 0

No indexed citations yet.

Article metrics

Real usage data collected on this platform.

38

Page views

0

PDF downloads

4

Outbound clicks

0

Citations

Views over time

Views by country

Approximate, from request IP at view time — not citizenship or institution. Countries with fewer than 5 views are grouped as "Other".

Traffic sources

Referring site, by host.

Views and downloads exclude known bots/crawlers. Citations combines this platform's own DOI-resolved index with each external source's own reported total — see Cited by above for individually listed citing works. Last refreshed 0 seconds ago.