Comparison of the Automatic Non-invasive Express Screening Analyser (ANESA)® for Clinical Analytical Parameters
Antoni Sicras-Mainar, Luis Collado-Yurrita
Journal of Advances in Medicine and Medical Research · pp. 1–17 · Published 10 Aug 2016
10.9734/BJMMR/2016/28216Abstract
Background: The objective we evaluated the reliability of some haematological and biochemical parameters performed by a non-invasive auto-analyser (ANESA) with those obtained by the standard method of venipuncture (reference method) in patients who went to the clinical analysis laboratory (Municipal-Hospital of Badalona, Spain). Methods: A transversal, comparative and parallel (paired) study was carried out. Two methods of study were practiced for the same subject: a) the reference method of venipuncture (conventional clinical analysis) and b) placement of sensors (comparison method: ANESA device). Consecutive patients older than 18 years, who met certain criteria for inclusion were included in the study during an 8 week period in 2014. The parameters studied were: haemogram (7), glucose, lipids (4), transaminases (2), bilirubin, creatinine and urea. Statistical analysis compared averages for paired groups and reliability of the obtained observations (method: intraclass-correlation coefficient (ICC); individual differences: Bland–Altman method), p<0.05. Results: A total of 195 patients were involved, with an average age of 50.8 years; 65.2% were women. In paired comparisons, cholesterol (185.4 vs. 179.8; difference: 5.6 mg/dL; p=0.005), cLDL (95.9 vs. 100.5; difference: −4.6 mg/dL; p=0.002) and bilirubin (0.6 vs. 0.5; difference: 0.1 mg/dL; p<0.001) obtained more modest results. Erythrocytes, haemoglobin, haematocrit, platelets, leukocytes, glucose, cHDL, triglycerides, ALT, AST, creatinine and urea reached an ICC>0.90. Lipid parameters (cholesterol: ICC=0.728; cLDL: ICC=0.817) obtained a moderate correlation, whereas lymphocytes (ICC=0.551) and monocytes (ICC=0.546) reached discrete results. Conclusions: Despite of the study limitations, the automatic non-invasive analyser (ANESA) is shown as a reliable and promising screening method in usual clinical practice for most of the analyzed parameters as an alternative to standard blood extraction. However, more studies are required to strengthen the consistency of the results.
Cited by 0
No indexed citations yet.
Related research
- Anxiety and Visual Field Assessment Reliability in Glaucoma Patients — shares topic coverage
- Home Recording And Video Selection: Their use In A Low-Resource Setting For Epilepsy Diagnosis — shares topic coverage
- A study on Clinical Validity and Reliability of the Bengali Language Translated Gastroesophageal Reflux Disease Quality of Life Questionnaire (GERD-QOL) in Gastroesophageal Reflux Disease Patients — shares topic coverage
- Reliability of the General Nutrition Knowledge Questionnaire among Head Teachers from Schools in Uganda — shares topic coverage
- Validity and Reliability of Combination of Observed Practice with Objective Structured Clinical Examination (OSCE) in Midwifery Final Exam — shares topic coverage
Article metrics
Real usage data collected on this platform.
0
Page views
0
PDF downloads
0
Outbound clicks
0
Citations
Views by country
Approximate, from request IP at view time — not citizenship or institution. Countries with fewer than 5 views are grouped as "Other".
No views recorded yet.
Traffic sources
Referring site, by host.
No traffic recorded yet.
Views and downloads exclude known bots/crawlers. Citations combines this platform's own DOI-resolved index with each external source's own reported total — see Cited by above for individually listed citing works. Last refreshed 0 seconds ago.